HCT-TT V1.3.6:
 1. [TTECL-462] : HCT-TT: BLE: Test case with measurement Mode Peak hold or First Peak will be returned with Ack 0x0b
 2. [TTECL-463] : HCT-TT: Default HID language
 3. [TTECL-464] : HCT-TT: Wrong overload value in "overload info" menu for 1st-Peak Mode

HCT-TT V1.3.5:
 1. [TTECL-462] : HCT-TT: BLE: Test case with measurement Mode Peak hold or First Peak will be returned with Ack 0x0b

HCT-TT V1.3.4:
 1. [TTECL-462] : HCT-TT: BLE: Test case with measurement Mode Peak hold or First Peak will be returned with Ack 0x0b

HCT-TT V1.3.3:
 1. [TTECL-461] : HCT-TT: In mode Peak Hold HCT, Sign of TW is sent via HID, even though unit has been turned off in HID settings
 2. [TTECL-457] : Calibrator2 SW: Overload Info: Export Overload Info (Excel) is not saving the correct overload info
 3. [TTECL-453] : HCT-TT Ext: Max and Min Values Limits in different units
 4. [TTECL-446] : HCT-TT: HCT-TT: Track HCT and Peak Hold HCT: Behavior of HCT-TT when TW connected is suddenly powered OFF
 5. [TTECL-440] : HCT-TT: Performing a new measurement by applying torque again not possible during HID Connection

HCT-TT V1.3.2:
 1. [TTECL-454] : HCT-TT-Ext: Implement lowest detection of 1st Peak according to existing TT sizes
 2. [TTECL-452] : HCT-TT: Track HCT and Peak Hold HCT: Advertising devices other than torque wrenches
 3. [TTECL-447] : HCT-TT: It is not possible to transfer a Testcase in "in.lb"
 4. [TTECL-427] : TT Extension (0,1-2Nm): Implementation of Unit cNm

HCT-TT V1.3.1:
 Automatically overwrite error parameters after startup (for 2Nm device)
 1. [TTECL-424] : TT Extension (0.1-2Nm): Target value can not be set
 2. [TTECL-425] : TT Extension (0.1-2Nm): Wrong target value shown on mainscreen
 3. [TTECL-427] : TT Extension (0,1-2Nm): Implementation of Unit cNm
 4. [TTECL-430] : TT-Extension: Resolution/Digits on Main Screen
 5. [TTECL-433] : HCT-TT BLE: Quick Connect is not working if Search for HCT-Devices is started in Connection Menue
 6. [TTECL-434] : T-Extension: Defualt values and overload values
 7. [TTECL-440] : HCT-TT: Performing a new measurement by applying torque again not possible during HID Connection
 8. [TTECL-441] : HCT-TT Extension (0,1-2Nm): Unit Resolution is not implemented as specified
 9. [TTECL-444] : HCT-TT Extension: Peak values are not displayed correctly
10. [TTECL-447] : HCT-TT: It is not possible to transfer a Testcase in "in.lb"
11. [TTECL-449] : HCT-TT: Unit cannot be transfered to TS
12. [TTECL-450] : HCT-TT Ext: Extend tolerance range
13. [TTECL-451] : HCT-TT Extension: Peak Hold HCT: HCT-TT does not recognize a compatible HCT-TS when Testcase is configured in "cNm" units

HCT-TT V1.3.0:
 1. [TTECL-442] : HCT-TT Extension: Peak Hold HCT and Track HCT: TS Live values are not displayed in these modes

HCT-TT V1.2.5:
 1. [TTECL-443] : Minimum response torque to the first peak.

HCT-TT V1.2.4:
 1. [TTECL-424] : TT Extension (0.1-2Nm): Target value can not be set
 2. [TTECL-425] : TT Extension (0.1-2Nm): Wrong target value shown on mainscreen
 3. [TTECL-427] : TT Extension (0,1-2Nm): Implementation of Unit cNm
 4. [TTECL-428] : TT Extension (0,1-2Nm): Deblocking process for HCT-TS
 5. [TTECL-429] : HCT-TT: Deblocking process with 12Nm TT for TS size 0.8-5Nm
 6. [TTECL-430] : TT-Extension: Resolution/Digits on Main Screen
 7. [TTECL-434] : TT-Extension: Defualt values and overload values

HCT-TT V1.2.3:
update proxy library to v1.5.0
 1. [TTECL-419] : HCT-TT BLE: Target value is manipulated when sending it in lbft or lbinch
 2. [TTECL-420] : HCT-TT BLE: Wrong internal calculation when sending Target value in lbinch or lbft
 3. [TTECL-424] : TT Extension (0.1-2Nm): Target value can not be set
 4. [TTECL-425] : TT Extension (0.1-2Nm): Wrong target value shown on mainscreen
 5. [TTECL-431] : HCT-TT: Change wording of "Threshold" to "Pre-Warning"

HCT-TT V1.2.2:
 1. [TTECL-418] : HCT-TT BLE: DeviceMeasurementControl.MeasurementJobId is -1, should be 0, also if new MeasurementCaseConfiguration with ID 0 is activated.
 2. [TTECL-419] : HCT-TT BLE: Target value is manipulated when sending it in lbft or lbinch
 3. [TTECL-420] : HCT-TT BLE: TT acknowledges HCT-Error 0x0B (Range) when sending Target value in lbinch
 4. [TTECL-421] : HCT-TT BLE: Target value is not rounded correctly
 5. [TTECL-423] : HCT-TT: Torque test case result information is not cleared when "Send" button is not pressed

HCT-TT V1.2.1:
 1. [TTECL-405] : HCT-TT: Introduction of Live Values Actual Torque and Peak Torque
 2. [TTECL-408] : HCT-TT: 1000Nm - Target value issue in combination with WinApp

HCT-TT V1.2.0:
 1. [TTECL-405] : HCT-TT: Introduction of Live Values Actual Torque and Peak Torque
 2. [TTECL-404] : HCT-TT: DTW Serial number not shown in CSV file
 3. [TTECL-403] : HCT-TT: "Configuration via App" screen according to translation table
 
HCT-TT V1.1.1:
 1. [TTECL-397] : HID: Configuration: HID Keyboard Layout remains unchanged
 2. [TTECL-398] : HID: Switching from HID to APP-Connection or vice versa does not work.
 3. [TTECL-400] : HCT-TT: Display of wrong target value when switching on and APP sends a new MeasurementCaseConfiguration
 4. [TTECL-401] : HID: When switching off HID and switching on afterwards, no transfer to PC anymore

HCT-TT V1.1.0:
 1. [TTECL-397] : HID: Configuration: HID Keyboard Layout remains unchanged
 2. [TTECL-398] : HID: Switching from HID to APP-Connection or vice versa does not work.
 3. [TTECL-399] : BLE Modul: Bluetooth Reset does not start advertising

HCT-TT V1.0.11:
 1. [TTECL-388] : Show test case name in Home screen up to 10 characters
 2. [TTECL-391] : BLE Modul: Configure HID Connection
 3. [TTECL-392] : HID: Behavior: If HID is activated, then only send HID-Data, wenn SEND Button is pressed
 4. [TTECL-393] : HID Configuration
 5. [TTECL-395] : HCT-TT: Blocked 500&850Nm DTW can not be deblocked with 1000Nm TT

HCT-TT V1.0.10:
 1. [TTECL-387] : HCT-TT BLE: Continue with new measurement after result

HCT-TT V1.0.9:
 1. [TTECL-386] : HCT-TT BLE: If TT is in result screen and HCT-Windows App wants to activate a new measurement mode, it is not possible
 2. [TTECL-387] : HCT-TT BLE: Continue with new measurement after result

HCT-TT V1.0.8:
 1. [TTECL-384] : HCT-TT: Read the current history entry by HCT-App

HCT-TT V1.0.7:
 1. [TTECL-384] : HCT-TT: Read the current history entry by HCT-App

HCT-TT V1.0.6:
 1. [TTECL-383] : HCT-TT: Issue when turning TT off/on with BLE Off
 2. [TTECL-382] : HCT-TT: TW Tolerance Min/Max not empty in Peak Hold and 1st Peak
 3. [TTECL-380] : HCT-TT: Wrong Measurement Status in CSV 
 4. [TTECL-213] : HCT-TT: Measurement not OK if you hit the edge of the tolerance range (low priority - do not implement now)

HCT-TT V1.0.5:
 1. [TTECL-380] : HCT-TT: Wrong Measurement Status in CSV 
 2. [TTECL-381] : HCT-TT: Add "Calibration status" column in .csv file

HCT-TT V1.0.4:
 1. [TTECL-239] : Deblocking process: Target values not correctly shown - Comma "," as digit separator
 2. [TTECL-371] : HCT-TT: Deblocking process - TT remains in "connecting" screen
 3. [TTECL-372] : HCT-TT: Deblocking process - Wrong Testcase is shwon after deblocking process canceled
 4. [TTECL-373] : HCT-TT: When bluetooth connection ends the screen "Configuration via Mobile App" should disappear 
 5. [TTECL-376] : HCT-TT: Track HCT Mode: Deblocking process interrupted. Strange behavior by connecting other devices
 6. [TTECL-377] : HCT-TT: History and Setting tab deactivated

HCT-TT V1.0.3:
 1. [TTECL-369] : HCT-TT: Frequent FW crashes while overloading TT
 2. [TTECL-366] : HCT-TT: Cancel/Go back start of Deblocking Process
 3. [TTECL-355] : HCT-TT: Enter Date&Time screen at the beginning
 4. [TTECL-213] : HCT-TT: Measurement not OK if you hit the edge of the tolerance range (low priority - do not implement now)

HCT-TT V1.0.2:
 1. [TTECL-356] : HCT-TT: Track HCT: Sporadic disconnection of torque wrench
 2. [TTECL-357] : HCT-TT: Quick Connect: Problem by connecting different TWs in a test case
 3. [TTECL-360] : HCT-TT: Quick Connect: Problem by choosing alternative test cases when TW does not fit
 4. [TTECL-361] : HCT-TT: Quick Connect: BLE: DeviceMeasurementCaseConfiguration is not published if TW does not fit and TT goes in Track HCT
 5. [TTECL-362] : HCT-TT: Mode is deleted when all Test Cases will be deleted. 
 6. [TTECL-363] : HCT-TT: Quick Connect: A test case that has already been deleted returns as a zombie.
 7. [TTECL-365] : HCT-TT: Connect Device: Wrench not fit to test case: Not possible to go back

HCT-TT V1.0.1:
 1. [TTECL-358] : HCT-TT: Quick Connect: Hang up If a TW does not fit to an already activated TestCase
 2. [TTECL-357] : HCT-TT: Quick Connect: Problem by connecting different TWs in a test case

HCT-TT V1.0.0:
 1. [TTECL-353] : HCT-TT: Menu>>Settings>>Display: Date and Time options are not corrected colored
 2. [TTECL-354] : HCT.TT: .csv File: Column W - "TT Serial Number" is not in the corrected column fulfilled.

HCT-TT V0.0.24:
 1. [TTECL-337] : HCT-TT: History.csv: TW Effective Length is shown times ten
 2. [TTECL-342] : HCT-TT: If already connected, optimize searching for devices
 3. [TTECL-344] : HCT-TT: Accelarate history read out
 4. [TTECL-349] : HCT-TT: Translation of new setting option "Set Date and Time" according translation table
 5. [TTECL-350] : HCT-TT: CSV File - Column F "TT Test Status" always "OK"

HCT-TT V0.0.23:
 1. [TTECL-337] : HCT-TT: History.csv: TW Effective Length is shown times ten
 2. [TTECL-338] : HCT-TT: Languages CSV file
 3. [TTECL-343] : HCT-TT: CSV File; Workflow Name & Active workflow position
 4. [TTECL-344] : HCT-TT: Accelarate history read out
 5. [TTECL-345] : HCT-TT: Date and Time configuration screen & additional setting via Menu
 6. [TTECL-346] : HCT-TT: Additional setting: "Set Date&Time: ON/OFF"
 7. [TTECL-347] : HCT-TT: Peak Hold HCT - Test case configuration wrong available devices to be connected

HCT-TT V0.0.22:
 1. [TTECL-265] : HCT-TT: Evaluation screens for all modes according to specification of histroy detail spec
 2. [TTECL-275] : HCT-TT: Track and Track HCT - Overload measurements (above 105%) must be saved in History and Overload Info list
 3. [TTECL-300] : HCT-TT: First Peak, Peak Hold and Peak Hold HCT: Measurement over 100% is not been reseted
 4. [TTECL-320] : HCT-TT: NO BLE Version - Test Mode displayed
 5. [TTECL-327] : HCT-TT: Connection procedure stops but still holds connection, next connection not possible
 6. [TTECL-331] : HCT-TT: Main screen flickers after "First Peak" mode is selected
 7. [TTECL-332] : HCT-TT: First Peak Target value NO: History entry wrong colored
 8. [TTECL-336] : HCT-TT: Peak Hold HCT and Track HCT - Change of mode - Target value range is limited
 9. [TTECL-337] : HCT-TT: History.csv: TW Effective Length is shown times ten
10. [TTECL-338] : HCT-TT: Languages CSV file
11. [TTECL-339] : HCT-TT: Torque Response Threshold (Torque started to be displayed)
12. [TTECL-340] : HCT-TT: Peak Hold HCT Mode: TT350 and TT1K: List of Devices: Any Torque wrench is available to be connected
13. [TTECL-341] : HCT-TT: Delete peak value of TW  "Peak:...." in top left corner 
14. [TTECL-342] : HCT-TT: If already connected, optimize searching for devices
15. [TTECL-343] : HCT-TT: CSV File; Workflow Name & Active workflow position

HCT-TT V0.0.21:
 1. [TTECL-265] : HCT-TT: Evaluation screens for all modes according to specification of histroy detail spec
 2. [TTECL-281] : HCT-TT: BLE: Despite APP-Connection is OFF TT is advertising
 3. [TTECL-297] : HCT-TT: Freeze in searching for devices with unusual signature on serial interface
 4. [TTECL-300] : HCT-TT: First Peak, Peak Hold and Peak Hold HCT: Measurement over 100% is not been reseted
 5. [TTECL-308] : HCT-TT: Track HCT and Peak Hold HCT modes: Try to establish a connection when Bluetooth is disabled
 6. [TTECL-318] : HCT-TT: after 1000 Measurements BLE symbol isn't blinking
 7. [TTECL-319] : HCT-TT: search function not working after entering quick connect
 8. [TTECL-321] : HCT-TT: Deblocking process successful already after first measurement
 9. [TTECL-324] : HCT-TT: TW remains connected after disconnecting
10. [TTECL-325] : HCT-TT: HCT TW gets disconnected if  menu-->connection-->"App"  is switched from ON to OFF
11. [TTECL-327] : HCT-TT: Connection procedure stops but still holds connection, next connection not possible
12. [TTECL-328] : HCT-TT: HCT TW main screen display error if menu-->connection-->"App" is switched from OFF to ON
13. [TTECL-329] : HCT-TT: History fast screen, History details screen: Values shown are cutted
14. [TTECL-330] : HCT-TT: Overload during deblocking causes hang-up in recalibration screen
15. [TTECL-331] : HCT-TT: Main screen flickers after "First Peak" mode is selected
16. [TTECL-332] : HCT-TT: First Peak Target value NO: History entry wrong colored
17. [TTECL-333] : HCT-TT: Trigger for BLE-Download does not work

HCT-TT V0.0.20:
 1. [TTECL-247] : HCT-TT: Peak Hold and First Peak with NO Target value: Not possible to navigate in the menu after a measurement
 2. [TTECL-260] : HCT-TT: Deblocking process: not possible to navigate backwards and exit deblocking process
 3. [TTECL-265] : HCT-TT: Evaluation screens for all modes according to specification of histroy detail spec
 3. [TTECL-267] : HCT-TT: NO-BLE TT Version
 4. [TTECL-275] : HCT-TT: Track and Track HCT - Overload measurements (above 105%) must be saved in History and Overload Info list
 5. [TTECL-277] : HCT-TT: issue after 1000 measurements in Peak hold without target
 6. [TTECL-281] : HCT-TT: BLE: Despite APP-Connection is OFF TT is advertising
 7. [TTECL-289] : HCT-TT: Quick Connect: Soft key does not work if TT-Send button is not pressed after measurement --> SW freezes
 8. [TTECL-294] : HCT-TT: After deletion of testcase, TT freezes in "save as Testcase" screen
 9. [TTECL-295] : HCT-TT: Peak Hold mode NO target value: Wrong mode configuration
10. [TTECL-296] : HCT-TT: Check for invalid parameters over BLE
11. [TTECL-297] : HCT-TT: Freeze in searching for devices with unusual signature on serial interface
12. [TTECL-299] : HCT-TT: Track and Track HCT: Measurement saved in History after a mode is changed
13. [TTECL-300] : HCT-TT: First Peak, Peak Hold and Peak Hold HCT: Measurement over 100% is not been reseted
14. [TTECL-301] : HCT-TT: First Peak and Peak Hold: HCT-TT freezed after button1 is pressed
15. [TTECL-302] : HCT-TT: First Peak and Peak Hold - Target value YES: Successfull measurement displayed in History 
16. [TTECL-303] : HCT-TT: Deblocking Process - TT remains in deblocking process after the checkmark of screen "Exit deblocking process?" is pressed  
17. [TTECL-304] : HCT-TT: Deblocking process - Color of evaluation is shown different during measurement and in evaluation screen
18. [TTECL-305] : HCT-TT: Deblocking process - After successful deblocking process, the mainscreen is displayed wrong
19. [TTECL-306] : HCT-TT: Deblocking process - Date and time configuration shown
20. [TTECL-307] : HCT-TT: First Peak and Peak Hold: History Detail entries not correctly displayed 
21. [TTECL-308] : HCT-TT: Track HCT and Peak Hold HCT modes: Try to establish a connection when Bluetooth is disabled
22. [TTECL-309] : HCT-TT: Peak Hold HCT, Target value NO, Tol TW NO: History Detail entry: Peak TT white colored in TT range
23. [TTECL-310] : HCT-TT: Peak Hold HCT, Target value NO, Tol TW YES: History Detail entry: Peak TT white colored in TT range
24. [TTECL-311] : HCT-TT: Peak Hold HCT, Target value YES, Tol TW YES: History Detail entry: Direction of rotation
25. [TTECL-312] : HCT-TT: List of devices is empty when a normal w/o HCT is activated


HCT-TT V0.0.19:
 1. [TTECL-144] : HCT-TT: "Tare not successful"  not shown
 2. [TTECL-178] : HCT-TT: negative torque value in history
 3. [TTECL-190] : HCT-TT: Changed wording of MIN/MAX tolerances
 4. [TTECL-198] : HCT-TT: Menu >> Settings >> Measurement >> Buzzer: OFF - Then buzzer is still activated at 100% even the option is deactivated.
 5. [TTECL-242] : HCT-TT: Date and Time configuration Screen
 6. [TTECL-246] : HCT-TT: Peak Hold and First Peak with NO Target value: Measured value must be always white colored
 7. [TTECL-247] : HCT-TT: Peak Hold and First Peak with NO Target value: Not possible to navigate in the menu after a measurement
 8. [TTECL-252] : HCT-TT: Torque visualization bar misbehavior is torque reaches exact upper tolerance value
 9. [TTECL-256] : HCT-TT: History-Detail Specification
10. [TTECL-258] : HCT-TT: Peak Hold HCT - Screenflow not correctly implemented: "Tol. Torque Wrench" screen not displayed when Target Value is NO
11. [TTECL-259] : HCT-TT: After deletion of testcase TT sends wrong Measurement_mode
12. [TTECL-260] : HCT-TT: Deblocking process: not possible to navigate backwards and exit deblocking process
13. [TTECL-262] : HCT-TT: Deblocking process - menu, List of devices etc. can be opened during deblocking process
14. [TTECL-263] : HCT-TT: Deblocking process -  wrong target values are used for the deblocking testcases
15. [TTECL-264] : HCT-TT: Deblocking process - no values of TW shown
16. [TTECL-265] : HCT-TT: Evaluation screens for all modes according to specification of histroy detail spec
17. [TTECL-267] : HCT-TT: NO-BLE TT Version
18. [TTECL-270] : HCT-TT: Headline doesn't get translated
19. [TTECL-271] : HCT-TT: Deblocking process - Applied torque during info screen is used for new/next measurement
20. [TTECL-273] : HCT-TT: "Overload" state with "History" screen behaviour
21. [TTECL-275] : HCT-TT: Track and Track HCT - Overload measurements (above 105%) must be saved in History and Overload Info list
22. [TTECL-277] : HCT-TT: issue after 1000 measurements in Peak hold without target
23. [TTECL-280] : HCT-TT: Response threshold of 2.5% is too small
24. [TTECL-281] : HCT-TT: BLE: Despite APP-Connection is OFF TT is advertising
25. [TTECL-289] : HCT-TT: Quick Connect: Soft key does not work if TT-Send button is not pressed after measurement 
26. [TTECL-290] : HCT:TT : Hang up in Peak Hold HCT, when not in calibrated area but Send Button is pressed
27. [TTECL-291] : Peak Hold, Peak Hold HCT and First Peak with NO Target value: Send button is pressed after measurement, FW is freezed, no reaction
28. [TTECL-292] : HCT-TT: List of advertising device shows only 6 devices
29. [TTECL-293] : HCT-TT: Quick Connect: Change Disconnecting to the beginning of the process

HCT-TT V0.0.18:
 1. [TTECL-122] : HCT-TT: BLE: Disconnect in list of HCT devices does not work
 2. [TTECL-231] : HCT-TT: wrong behaviour when requesting more entrys than saved
 3. [TTECL-237] : Deblocking process: "Connection via Menu>>Connection>>HCT Devices" not possible
 4. [TTECL-239] : Deblocking process: Target values not correctly shown
 5. [TTECL-241] : Connecting screen flickers
 6. [TTECL-242] : HCT-TT: Date and Time configuration Screen
 7. [TTECL-243] : HCT-TT: Peak Hold HCT - Wrong behaviour when connecting to TW without target value
 8. [TTECL-244] : HCT-TT: All modes - Reverse Screenflow for "Traget value"  not correct 
 9. [TTECL-246] : HCT-TT: Peak Hold and First Peak with NO Target value: Measured value must be always white colored
10. [TTECL-247] : HCT-TT: Peak Hold and First Peak with NO Target value: Not possible to navigate in the menu after a measurement
11. [TTECL-248] : HCT-TT: Peak Hold and First Peak with NO Target value: Evaluation screen is shown if send butten is pressed during Evaluation timeout.
12. [TTECL-249] : HCT-TT: Track HCT mode: Direction of rotation symbol is not being updated
13. [TTECL-250] : HCT-TT: Track and Track HCT mode: Measurement saved in History
14. [TTECL-251] : HCT-TT: History Entry: Mode must be shown in the first lines
15. [TTECL-253] : HCT-TT: In Peak Hold HCT mode if unit in.lb or ft-lb is selected, the result is always evaluated as not ok
16. [TTECL-254] : HCT-TT: BLE: No Live values from HCT-TW due to missing activation of notification
17. [TTECL-255] : HCT-TT BLE: PeakHoldHCT without Target Value, DeviceMeasurementResult not correct
18. [TTECL-257] : HCT-TT: Modes with "no target value" - option do also evaluate if peak-value was not in the calibrated area
19. [TTECL-266] : HCT-TT: Peak Hold HCT: "Connect Device" screen, not possible to navigate backwards
20. [TTECL-272] : HCT-TT: Deblocking process - TT remains in deblocking process after TW is disconnected
21. [TTECL-273] : HCT-TT: "Overload" state with "History" screen behaviour
22. [TTECL-274] : HCT-TT Menu>>Settings>>System: Overload Info - Overloaded measurement is not saved in Overload Info
23. [TTECL-275] : HCT-TT: Track and Track HCT - Overload measurements (above 105%) must be saved in History and Overload Info list
24. [TTECL-276] : HCT-TT: Quick Connect: No "Connecting" Screen if device has been chosen  
25. [TTECL-278] : HCT-TT: BLE: Quick Connect does not work if already connected to a device
26. [TTECL-281] : HCT-TT: BLE: Despite APP-Connection is OFF TT is advertising
27. [TTECL-282] : HCT-TT: BLE: Take over corrections from the Garant Interface into ble_proxy_library
28. [TTECL-284] : HCT-TT: BLE: TT stucks in "Searching for HCT devices" if more than 5 Devices are advertising
29. [TTECL-286] : HCT-TT: PeakHold HCT with NO Target value: If send button is pressed after Measurement "No HCT-Device connected"
30. [TTECL-288] : HCT-TT: Wrong Menu symbol shown

HCT-TT V0.0.17:
 1. [TTECL-239] : Deblocking process: Target values not correctly shown
 2. [TTECL-232] : HCT-TT: Symbols during disconnecting not shown correctly
 3. [TTECL-227] : HCT-TT: Take over and activate DeviceMeasurementCaseConfiguration from APP
 4. [TTECL-226] : HCT-TT: Second value in history is not displayed correctly
 5. [TTECL-216] : HCT-TT: Track HCT and Peak Hold HCT: "not calibrated" message visualization also in DTW
 6. [TTECL-207] : HCT-TT: Peak Hold and 1st Peak: It is possible to set Min and Max Tolerance to 0,0% by editing a Test Case.
 7. [TTECL-201] : HCT-TT: Dimming Time Behavior and Torque applied implementation
 8. [TTECL-196] : HCT-TT: Track HCT Mode: Measured values displayed above 115% of nominal value muss be as "----" displayed
 9. [TTECL-191] : HCT-TT: Relaunch of the Peak-Tolerance/ TW-Tolerance
10. [TTECL-189] : HCT-TT: 1st Peak mode - Target value Yes/No
11. [TTECL-187] : HCT-TT: Live values in Track mode
12. [TTECL-186] : HCT-TT: EEPROM corrupt with 1001 measurements
13. [TTECL-175] : HCT-TT: Quick Connect shows a device even when trying to exit

HCT-TT V0.0.16:
 1. [TTECL-178] : HCT-TT: negative torque value in history
 2. [TTECL-182] : HCT-TT: DeviceAccessSyncronization with flag 1 not sent in QuickHistory/Connecting screens
 3. [TTECL-187] : HCT-TT: Live values in Track mode
 4. [TTECL-188] : HCT-TT: Live Values in Track HCT mode
 5. [TTECL-189] : HCT-TT: 1st Peak mode - Target value Yes/No
 6. [TTECL-191] : HCT-TT: Relaunch of the Peak-Tolerance/ TW-Tolerance
 7. [TTECL-192] : HCT-TT: Peak Hold & Peak Hold HCT  - Target value Yes/No
 8. [TTECL-196] : HCT-TT: Track HCT Mode: Measured values displayed above 115% of nominal value muss be as "----" displayed
 9. [TTECL-202] : HCT-TT: Alarm when Torque exceeds 125% of maximal load
10. [TTECL-207] : HCT-TT: Peak Hold and 1st Peak: It is possible to set Min and Max Tolerance to 0,0% by editing a Test Case.
11. [TTECL-216] : HCT-TT: Track HCT and Peak Hold HCT: "not calibrated" message visualization also in DTW
12. [TTECL-219] : HCT-TT: All modes: Measured values displayed above 115% of nominal value muss be as "----" displayed.
13. [TTECL-221] : HCT-TT: setting measurement_mode via BLE
14. [TTECL-222] : HCT-TT: Peak Hold HCT  - Target value Yes/No
15. [TTECL-226] : HCT-TT: Second value in history is not displayed correctly
16. [TTECL-227] : HCT-TT: Take over and activate DeviceMeasurementCaseConfiguration from APP
17. [TTECL-229] : HCT-TT: Wait for Connectionstate with Subscription for HCT (BLE Version 1.3.2)
18. [TTECL-230] : HCT-TT: measurement_result peak torque is incorrect
19. [TTECL-232] : HCT-TT: Symbols during disconnecting not shown correctly
20. [TTECL-235] : HCT-TT: Peak Hold HCT - History entries not correctly shown in history screen
21. [TTECL-236] : HCT-TT: Set Tolerance values not saved
22. [TTECL-238] : Deblocking process: "Deblocking successful" notification does not appear after a correct deblocking process
23. [TTECL-239] : Deblocking process: Target values not correctly shown
24. [TTECL-240] : Deblocking process: Torque during info screens is used in new measurements

HCT-TT V0.0.15:
 1. [TTECL-119] : HCT-TT: Read History via RACP needs to support partial read
 2. [TTECL-122] : HCT-TT: Disconnect in list of HCT devices does not work
 3. [TTECL-144] : HCT-TT: "Tare not successful"  not shown
 4. [TTECL-146] : HCT-TT: Add String IDs for language translation 
 5. [TTECL-159] : HCT-TT: Direction of rotation is always Clockwise
 6. [TTECL-175] : HCT-TT: Quick Connect shows a device even when trying to exit
 7. [TTECL-177] : HCT-TT: 350 TT test case values are always negative
 8. [TTECL-182] : HCT-TT: DeviceAccessSyncronization with flag 1 not sent in QuickHistory/Connecting screens
 9. [TTECL-183] : HCT-TT: Menu >> Connection >> HCT Devices: Other devices beside Garant DTW and Holex HCT-TW are discoverable
10. [TTECL-184] : HCT-TT: List of HCT Devices: Connection option despite any devices are found
11. [TTECL-185] : HCT-TT: Quick connect still shows already connected TW despite list query returns no devices
12. [TTECL-186] : HCT-TT: EEPROM corrupt with 1001 measurements
13. [TTECL-190] : HCT-TT: Changed wording of MIN/MAX tolerances
14. [TTECL-194] : HCT-TT: Tolerance for Deblocking process
15. [TTECL-198] : HCT-TT: Menu >> Settings >> Measurement >> Buzzer: OFF - Then buzzer is still activated at 100% even the option is deactivated.
16. [TTECL-199] : HCT-TT: Track Mode: Direction of rotation symbol is not visualized and updated during a measurement
17. [TTECL-200] : HCT-TT: Peak Hold: Test case with counter clockwise rotation not possible to be configured.
18. [TTECL-201] : HCT-TT: Dimming Time Behavior and Torque applied implementation
19. [TTECL-202] : HCT-TT: Alarm when Torque exceeds 125% of maximal load
20. [TTECL-205] : HCT-TT: Weird characters shown in password prompt
21. [TTECL-206] : HCT-TT: amount of numbers above bar chart - XX,XX/ XX,XXNm
22. [TTECL-208] : HCT-TT: Track HCT and Peak Hold HCT: Values recieved from DTW must be displayed without the sign during a measurement.
23. [TTECL-210] : HCT-TT: Target value for 1000Nm Version 
24. [TTECL-212] : HCT-TT: LIST OF HCT DEVICES: Connection Symbol must always colored green
25. [TTECL-217] : HCT-TT: Track HCT Mode: Direction of rotation symbol is not visualized and updated during a measurement

HCT-TT V0.0.14:
 1. [TTECL-23]  : HCT-TT: History & "Quick"-History
 2. [TTECL-116] : HCT-TT: Warning message when no TW connected and torque is applied
 3. [TTECL-118] : Delete all entries in history list via RACP
 4. [TTECL-119] : HCT-TT: Read History via RACP needs to support partial read
 5. [TTECL-123] : HCT-TT: If App-Connection is deactivated, then the TT does not need to start Advertising
 6. [TTECL-129] : HCT-TT: BLE: Quick Connect does not work due to no communication
 7. [TTECL-144] : HCT-TT: "Tare not successful"  not shown
 8. [TTECL-147] : HCT-TT: Implementation of language translations
 9. [TTECL-149] : If measured torque exceeds 125% of MaxTorque, the buzzer shall peep despite of the settings 
10. [TTECL-150] : HCT-TT: Date and Time not saved when power supply is interrupted
11. [TTECL-151] : HCT-TT: TT saves two results for one measurement
12. [TTECL-152] : HCT-TT: Second item in history shows only one measurement.
13. [TTECL-153] : HCT-TT: Disconnect option not working
14. [TTECL-154] : HCT-TT: Unknown password after TT firmware upgrade to v.0.0.13
15. [TTECL-155] : HCT-TT: Connection to mobile app will be lost once an HCT-Mode is activated
16. [TTECL-159] : HCT-TT: Direction of rotation is always Clockwise
17. [TTECL-161] : HCT-TT: Battery status not updated 
18. [TTECL-165] : HCT-TT: 350TT minium limit is wrong
19. [TTECL-166] : HCT-TT: No Buzzer sound when max. torque applied and Buzzer=OFF
20. [TTECL-167] : HCT-TT: Wrong wording of "connecting..." screen
21. [TTECL-168] : HCT-TT: Wording issue "screw case"
22. [TTECL-172] : HCT-TT: Ble Version returns 0.0.0 (must be updated to the corresponding version)
23. [TTECL-173] : HCT-TT: Next Calibration Date returns always the same value
24. [TTECL-175] : HCT-TT: Quick Connect shows a device even when trying to exit
25. [TTECL-176] : HCT-TT: Exiting Quick Connect multiple times causes the TT to freeze


HCT-TT V0.0.13:
 1. [TTECL-75]  : HCT-TT: 1st Peak Mode functionality
 2. [TTECL-83]  : HCT-TT: Disconnection of a connected torque wrench 
 3. [TTECL-93]  : HCT-TT: Wrong warning when in.lb is activated
 4. [TTECL-99]  : HCT-TT: RACP false number of history measurements
 5. [TTECL-115] : HCT-TT: Screen flickers
 6. [TTECL-116] : HCT-TT: Warning message when no TW connected and torque is applied
 7. [TTECL-118] : Delete all entries in history list via RACP
 8. [TTECL-119] : HCT-TT: Read History via RACP needs to support partial read
 9. [TTECL-120] : HCT-TT: Freezes if connection is lost
10. [TTECL-121] : Integration proxy-library V1.2.22
11. [TTECL-122] : HCT-TT: Disconnect in list of HCT devices does not work
12. [TTECL-123] : HCT-TT: If App-Connection is deactivated, then the TT does not need to start Advertising
13. [TTECL-125] : HCT-TT: Correct SGTIN in Scan response data
14. [TTECL-126] : HCT-TT: Unit Change is not transfered to an already connected TW
15. [TTECL-127] : HCT-TT: Missing checkmark for "Next calibration"
16. [TTECL-128] : HCT-TT: Compare of TW's peak value with his range value fails, if unit is different to Nm
17. [TTECL-129] : HCT-TT: BLE: Quick Connect does not work due to no communication
18. [TTECL-133] : HCT-TT: Input of target value not possible when unit is in.lb
19. [TTECL-134] : HCT-TT: Change wording for Quick connect screen
20. [TTECL-135] : HCT-TT: Do not show "Peak Tolerance" in History details
21. [TTECL-136] : HCT-TT: Mode order wrong in Track HCT
22. [TTECL-138] : HCT-TT: Warning recalibration symbol color
23. [TTECL-139] : HCT-TT: Pixel artifacts while making a measurement
24. [TTECL-140] : HCT-TT: Device access synchronisation.Tool config active is not set when entering Mode
25. [TTECL-141] : HCT_TT: DeviceMeasurementResult yet incomplete when HCT-Mode is activated
26. [TTECL-143] : HCT-TT: Able to exit Overload state
27. [TTECL-144] : HCT-TT: "Tare not successful"  not shown
28. [TTECL-148] : HCT-TT: Max. Torque not shown in red for Track & Track HCT 

HCT-TT V0.0.12:
 1. [TTECL-10]  : HCT-TT: Saving last set values of mode menu
 2. [TTECL-52]  : HCT-TT: BLE stops communicating
 3. [TTECL-68]  : Peak Hold HCT: result screen colors incorrect
 4. [TTECL-83]  : HCT-TT: Disconnection of a connected torque wrench 
 5. [TTECL-89]  : HCT-TT: Reset process of TT and TW after Measurement takes too long
 6. [TTECL-90]  : HCT-TT: Restart Testcases automatically when torque is applied 
 7. [TTECL-92]  : HCT-TT: Measurement unit no set correctly
 8. [TTECL-94]  : Test Case name remains after changing mode and factory reset
 9. [TTECL-95]  : Mode limits out of TW range
10. [TTECL-96]  : "Searching for HCT Devices" and "Connecting" only in quick menu
11. [TTECL-98]  : TT unblocking not allowed
12. [TTECL-99]  : RACP false number of history measurements
13. [TTECL-100] : BLE-Interface: TT application does not wait for the acknowledge when connected to the TW
14. [TTECL-101] : HCT-TT: BT ON<-->OFF too much delay
15. [TTECL-102] : HCT-TT: Missing checkmark for "HCT device" option 
16. [TTECL-103] : HCT-TT: Checkmark at "Quick connect" not needed
17. [TTECL-104] : HCT-TT: Delete "HCT-Master" setting option
18. [TTECL-105] : HCT-TT: Wording of "HCT device" setting
19. [TTECL-106] : HCT-TT: Sequence of setting options
20. [TTECL-107] : HCT-TT: BT Symbol red when no TW connected 
21. [TTECL-108] : HCT-TT: Max. applied torque value should be stay at mainscreen
22. [TTECL-109] : HCT-TT: Only show set test case name on mainscreen w/o "testcase:"
23. [TTECL-110] : HCT-TT: Overload not showed in history screen
24. [TTECL-111] : HCT-TT: Scrollbar in menu is being cut at the end
25. [TTECL-112] : HCT-TT: "BT zurücksetzen" wording
26. [TTECL-113] : HCT-TT: 1st Peak mode wrong wording
27. [TTECL-114] : HCT-TT: Peak tolerance 
28. [TTECL-116] : HCT-TT: Warning message when no TW connected and torque is applied
29. [TTECL-117] : HCT-TT: Changed wording "No HCT device connected"

HCT-TT V0.0.11:
 1. [TTECL-63] : HCT-TT: Wrong mode is shown on main screen
 2. [TTECL-65] : HCT-TT: Buzzer not working for threshold and target value indicating
 3. [TTECL-75] : HCT-TT: 1st Peak Mode do not start evaluation after 1st Peak
 4. [TTECL-81] : HCT-TT: No disconnect when choosing Track-Mode
 5. [TTECL-86] : HCT-TT: Connecting Screen during connection with TW
 6. [TTECL-87] : HCT-TT: Delete test case wording
 7. [TTECL-88] : HCT-TT: Info Screen: Searching for HCT devices...
 8. [TTECL-89] : HCT-TT: Reset process of TT and TW after Measurement takes too long

HCT-TT V0.0.10:
 1. [TTECL-23] : HCT-TT: History & "Quick"-History
 2. [TTECL-30] : Initialize HCT Measurement Process
 3. [TTECL-33] : Check and filtering of Test cases with TW Nm ranges
 4. [TTECL-48] : History via RACP mechanism
 5. [TTECL-55] : HCT-TT: Modes order while creating a test case changes
 6. [TTECL-57] : HCT-TT: Unit selection not showing properly
 7. [TTECL-58] : HCT-TT: Deblocking process
 8. [TTECL-61] : HCT-TT: Change wording from "Screw case PWD" to "Test case PWD"
 9. [TTECL-62] : HCT-TT: "HCT" not shown for HCT Test cases 
10. [TTECL-63] : HCT-TT: Wrong mode is shown on main screen
11. [TTECL-64] : HCT-TT: Recalibration and Blocked information/alert
12. [TTECL-65] : HCT-TT: Buzzer not working for threshold and target value indicating
13. [TTECL-67] : HCT-TT: Connection takes too long
14. [TTECL-68] : Peak Hold HCT: result screen colors incorrect
15. [TTECL-69] : Protocol Type Default Value Incorrect
16. [TTECL-70] : Enabled factors show Vibration is active when the TT doesn't have one.
17. [TTECL-71] : HCT-TT: Active Workflow Position Default Value Incorrect
18. [TTECL-72] : HCT-TT: Measurement Units Default Value Incorrect 
19. [TTECL-74] : HCT-TT: Settings--> System --> "Screw case PW" ; Wrong wording
20. [TTECL-75] : HCT-TT: 1st Peak Mode do not start evaluation after 1st Peak
21. [TTECL-76] : HCT-TT; Target value should be shown on main screen
22. [TTECL-77] : HCT-TT: Naming of a Test Case only with numbers  
23. [TTECL-79] : HCT-TT: Info message when no torque wrench is connected
24. [TTECL-80] : Test Case that doesn't fit connects to incorrect selected wrench
25. [TTECL-81] : HCT-TT: No disconnect when choosing Track-Mode

HCT-TT V0.0.9:
 1. [TTECL-53] : Bluetooth does not show devices

HCT-TT V0.0.8:
update proxy lib to ble_proxy_win_V1.2.19
 1. [TTECL-34] : HCT-TT: First Peak Mode: Peak value is refreshed
 2. [TTECL-37] : HCT-TT: App connection ON/OFF
 3. [TTECL-40] : Quick Connect
 4. [TTECL-41] : SW-Trigger for FW-Update over the air.
 5. [TTECL-42] : HCT-TT: First Peak mode not available
 6. [TTECL-43] : HCT-TT: Menu >> Connection: Quick Connect should only be editable when HCT Master is ON

HCT-TT V0.0.7:
update proxy lib to ble_proxy_win_V1.2.16
 1. [TTECL-9] : HCT-TT: Wrong serialnumber shown in mobile App
 2. [TTECL-11] : HCT-TT: Test Cases remains after creating a mode
 3. [TTECL-12] : HCT-TT: minimum adjustable Torque for 350Nm version is 70Nm instead of 20Nm
 4. [TTECL-13] : HCT-TT: HCT-TT: Last and Next Calibration date 
 5. [TTECL-14] : HCT-TT: Device name missing at Info-Screen
 6. [TTECL-15] : HCT-TT: Switch automatically back to main screen after connecting a torque wrench.
 7. [TTECL-16] : HCT-TT: Peak values of TW and TT on main screen
 8. [TTECL-19] : HCT-TT: Greyed out settings in connection menu
 9. [TTECL-20] : HCT-TT: Color change of selcted history value 
10. [TTECL-21] : HCT-TT: Test case editing --> "Test case existing"
11. [TTECL-22] : HCT-TT: Color change during history navigation
12. [TTECL-23] : HCT-TT: History & "Quick"-History
13. [TTECL-24] : HCT-TT: Function of left softkey changes during track mode
14. [TTECL-25] : HCT-TT: Sequence of modes navgation not identical
15. [TTECL-29] : Measurement Modes and Splitscreen
16. [TTECL-31] : Check and filtering of TT with TW Nm ranges
17. [TTECL-32] : Measurement and Synchronisation between TW and TT
18. [TTECL-33] : Check and filtering of Test cases with TW Nm ranges
19. [TTECL-35] : HCT-TT: Peak Hold/First Peak modes: HCT-TT freezed after not succesfully measurement
20. [TTECL-36] : HCT-TT: Information for not calibrated range
21. [TTECL-38] : HCT-TT: Changed local name for advertisment
22. [TTECL-39] : HCT-TT: Peak value of TW remains on main screen

HCT-TT V0.0.6:
Update proxy lib to ble_proxy_win_V1.2.13

HCT-TT V0.0.5:
BLE Slave functionality
BLE Master functionality
Update proxy lib to ble_proxy_win_V1.2.12

HCT-TT V0.0.4:
TT Screens_PA02 ""                        : ID3,ID6,ID7,ID8,ID9,
TT Screens_PA02 "Track- Mode"             : ID13
TT Screens_PA02 "1st Peak- Mode"          : ID20,ID21,ID22,ID23,ID24,ID25,ID26
TT Screens_PA02 "Peak Hold"               : ID32
TT Screens_PA02 "1st Peak HCT-Master ON"  : ID34,ID37
TT Screens_PA02 "Peak Hold HCT-Master ON" : ID42,ID43
TT Screens_PA02 "History"                 : ID45,ID46
TT Screens_PA02 "Test Case"               : ID49,ID50,ID51,ID52,ID53,ID54
TT Screens_PA02 "Connection"              : ID57
TT Screens_PA02 "Settings"                : ID67,ID68
TT Screens_PA02 "Info"                    : ID70